Measurement of circulating 25-hydroxy vitamin D using three commercial enzyme linked immunosorbent assay kits with comparison to liquid chromatography - tandem mass spectrometry method.
journal contribution
posted on 2014-01-08, 14:01 authored by Cheng-Shiun He, Michael Gleeson, William D. FraserThe purpose of this study was to compare the accuracy and clinical implications of three commercial enzyme-linked immunosorbent assay (ELISA) kits (Eagle Biosciences, Immundiagnostik, and MicroVue) with a validated liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the measurement of serum 25(OH)D concentration. Methods. Blood samples were obtained from 225 healthy individuals who were recruited as subjects from Loughborough University, UK. Plasma samples were measured for 25(OH)D concentration by means of LC-MS/MS and ELISA kits from Eagle Biosciences, Immundiagnostik, and MicroVue. Results. The 25(OH)D concentration measured by the Eagle Biosciences, Immundiagnostik, and MicroVue ELISAs biased −50.9 ± 79.1 nmol/L, −14.2 ± 91.0 nmol/L, and −7.2 ± 18.9 nmol/L (bias ± SD) from the LC-MS/MS method, respectively. We found that 52% (Eagle Biosciences), 48% (Immundiagnostik), and 38% (MicroVue) of participants were misclassified, and the results showed the poor agreement (Kappa: −0.201~0.251) in classification of participants defined as vitamin D sufficiency and insufficiency between each method and LC-MS/MS. Conclusions. The present study demonstrated that there were negative biases and considerable misclassification of participants using the cut-off point (50 nmol/L) for vitamin D insufficiency and sufficiency using the Eagle Biosciences, Immundiagnostik, and MicroVue ELISAs compared with the LC-MS/MS assay.
History
School
- Sport, Exercise and Health Sciences
Citation
HE, C-S., GLEESON, M. and FRASER, W.D., 2013. Measurement of circulating 25-hydroxy vitamin D using three commercial enzyme linked immunosorbent assay kits with comparison to liquid chromatography - tandem mass spectrometry method. ISRN Nutrition, 2013, Article ID 723139.Publisher
Hindawi Publishing Corporation / © Cheng-Shiun He et al.Version
- VoR (Version of Record)
Publication date
2013Notes
This article was published in the journal ISRN Nutrition, it is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.ISSN
2314-4068Publisher version
Language
- en