posted on 2014-01-28, 15:32authored byJiawei Fu, Pablo RuizPablo Ruiz, Fabrice Pierron
In this project, the depth-resolved full-field deformation of the porcine cornea under changing intraocular
pressure was investigated by performing digital volume correlation (DVC) on the reconstructed volume images generated
through swept source optical coherence tomography (SS-OCT). Posterior inflation test of porcine cornea sample for two
load steps were performed and the distribution patterns of displacement and strain fields were produced. The error
sources for the measurements were analyzed. The refraction induced OCT image distortion is a main error source for the
measurement results. Then, a methodology was developed to correct the OCT distortion based on the Fermat’s principle.
History
School
Mechanical, Electrical and Manufacturing Engineering
Citation
FU, J., RUIZ, P.D. and PIERRON, F., 2013. Correction of refraction induced distortion in optical coherence tomography corneal reconstructions for volume deformation measurements. Presented at: Photomechanics 2013, 27th-29th May 2013, Montpellier, France.