posted on 2018-10-29, 11:01authored byAngelos S. Echiadis
Photoplethysmography, a technique to measure by optical means volume changes,
has been known and applied for many years. One of its most popular applications is
pulse oximetry, a non-invasive method to measure oxygen content in arterial blood.
It is based on the principle of arterial blood volume changes due to heart
contractions, known as systoles. Systolic pulsations appear on the arterial vascular
system, while blood flow in veins does not normally present pulsations, especially
at remote parts of the peripheral vascular system, such as the fingers. Therefore,
pulse oximetry is only applicable to arteries as their pulsations allow for separation
of the pulsatile components from the rest of the absorbing components. A novel
non-invasive technique permits the measurement of venous oxygen saturation by
introducing a series of pulsations in the veins thus allowing the separation of venous
signal components for calculation of venous oxygen saturation.
This thesis presents a theoretical model describing the mechanical coupling of
arteries and veins and its effects in the accuracy of oxygen saturation measurement. [Continues.]
Funding
Loughborough University, Department of Electronic and Electrical Engineering (studentship). BTG plc.
History
School
Mechanical, Electrical and Manufacturing Engineering
This work is made available according to the conditions of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) licence. Full details of this licence are available at: https://creativecommons.org/licenses/by-nc-nd/4.0/
Publication date
2007
Notes
A Doctoral Thesis. Submitted in partial fulfilment of the requirements for the award of Doctor of Philosophy at Loughborough University.