posted on 2018-06-28, 10:04authored byEugene R. Arulmuthu
The aim of this research has been to study and develop the engineering principles
associated with the impact of formulation and device parameters on the safe delivery
of nano-sized biomaterials such as plasmid DNA. In the present investigation, Omron
U22 and U03 mesh nebulisers operating at frequencies of ~175 kHz and ~65 kHz
respectively were used. Since the U22 device is a recently introduced mesh nebuliser
for respiratory drug delivery, detailed characterisation, experimentation, modelling and
analysis was carried out for this device.
Plasmids of size 5.7, 8.7, 13 and 20 kb were purified from Escherichia coli cells and
used for nebulisation experiments. Experiments on the nebulisation of plasmid DNA
using the U22 device in a bio-safety cabinet showed no damage to the sc structure of
the 5.7 kb plasmid, but almost complete damage to the 20 kb plasmid in the condensed
aerosols collected using a fabricated aerosol collection apparatus. The damage to the
sc structure of plasmid DNA was analysed using gel electrophoresis, PicoGreen assay
and atomic force microscope (AFM). Engineering analysis was performed using
computational fluid dynamics (CFD) modelling to determine the shear and elongational
strain rates in the mesh nozzle of nebuliser. The estimated maximum hydrodynamic
force on plasmid DNA based on the Ryskin equation was calculated in picoNewton
(PN) from the actual molecular size of the sc structure and predicted strain rates.
Optimisation of the formulation and device parameters were carried out using Design
of Experiments (DOE) to predict damage to the sc structure. Formulation of the 20 kb
plasmid with polyethyleneimine (PEI) resulted in safe aerosol delivery using the mesh
nebuliser. In vitro transfection studies in suspension-adapted Chinese Hamster Ovary
(CHO-S) cells resulted in successful integration of Green Fluorescent Protein (GFP)
from the 5.7 kb plasmid after nebulisation.
The commercially available U22 mesh nebuliser promises to be a useful pulmonary
device for the successful delivery of plasmid DNA for non-viral gene therapy.
Realisation of this promise however will require both innovations in the design of
experiments, formulation and methods of studying plasmid DNA damage as
demonstrated in this thesis.
History
School
Mechanical, Electrical and Manufacturing Engineering
This work is made available according to the conditions of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) licence. Full details of this licence are available at: https://creativecommons.org/licenses/by-nc-nd/4.0/
Publication date
2007
Notes
A Doctoral Thesis. Submitted in partial fulfilment of the requirements for the award of Doctor of Philosophy at Loughborough University.